source: FOIAVistA/tag/r/AUTOMATED_LAB_INSTRUMENTS-LA/LAMIV12.m@ 628

Last change on this file since 628 was 628, checked in by George Lilly, 14 years ago

initial load of FOIAVistA 6/30/08 version

File size: 1.4 KB
Line 
1LAMIV12 ;SLC/DLG - PROCESS VITEK GNS CARDS ;7/20/90 09:38 ;
2 ;;5.2;AUTOMATED LAB INSTRUMENTS;**5**;Sep 27, 1994
3A Q:$E(IN,1,2)'="01" ;NO RESULTS OR MULTI COLUMN
4 S RMK=$E(IN,3),RMK=$S(RMK=0:"Oxidase Negative",RMK=1:"Oxidase Positive",1:"")
5 S O=$E(IN,5,6),V=$E(IN,7),TYPE=$S("12"[V:$O(^LAB(62.4,TSK,7,"B","0E",0)),1:"") Q:TYPE<1
6 S V=$O(^LAB(62.4,TSK,7,TYPE,1,"C",O,0)) Q:V<1 S ORG(ISOL)=+^LAB(62.4,TSK,7,TYPE,1,V,0)
7 F I2=1:1:13 S V=$E(IN,(I2*6)+11),I3=$E(IN,16) D MIC0
8 Q
9MIC0 S O=$E(IN,(I2*6)+15),I4=$O(^LAB(62.4,TSK,7,CARD,2,"C",O,0)) Q:I4'>0 S I3=$P(^LAB(62.4,TSK,7,CARD,2,I4,0),U,2) X $P(^(0),U,3) S:V]"" MIC(ISOL,I3)=V
10 Q
11512 S C=0,U="^" ;0E, 5.12, Gram Neg ID card
12 F I=3,7 S X1=$O(^LAB(62.4,TSK,7,CARD,1,"C",$E(IN,I,I+1),0)) D L2:X1>0 Q:$D(ORG)
13 S RMK="",CODE=43,FL=$E(IN,44) D MSG S CODE=44,FL=$E(IN,44) D MSG S CODE=45,FL=$E(IN,46) D MSG S CODE=46,FL=$E(IN,47) D MSG S CODE=47,FL=$E(IN,48) D MSG
14 S:'$D(ORG(ISOL)) BAD=1 Q
15L2 S ORG(ISOL)=+^LAB(62.4,TSK,7,CARD,1,X1,0),X=$E(IN,I+2,I+3) D PROB S ORG(ISOL,1)=X I X=""!(X<80) K ORG(ISOL) ;cancel if PROB<80%
16 Q
17HEX S XX=X,X="" F II=1:1:$L(XX) S X=X*16+($F("0123456789ABCDEF",$E(XX,II))-2)
18 Q
19MSG F X1=0:0 S X1=$O(^LAB(62.4,TSK,7,CARD,4,"B",CODE,X1)) Q:X1'>0 D MS2
20 Q
21MS2 S X3=^LAB(62.4,TSK,7,CARD,4,X1,0)
22 S X4=$P(X3,U,2) I $L(X4),X4'=FL Q ;
23 S:$L(RMK) RMK=RMK_", " S RMK=RMK_$P(X3,U,3)
24 Q
25PROB D HEX I X>100 S X="" Q
26 S:X=0 X="<1" S X=X_"% Probability" Q
Note: See TracBrowser for help on using the repository browser.