source: FOIAVistA/trunk/r/AUTOMATED_LAB_INSTRUMENTS-LA/LAH717U.m@ 1328

Last change on this file since 1328 was 628, checked in by George Lilly, 15 years ago

initial load of FOIAVistA 6/30/08 version

File size: 1.5 KB
Line 
1LAH717U ;SLC/DLG - HITACHI 717 ROUTINE FOR AUTOMATED DATA ;7/20/90 09:10 ;
2 ;;5.2;AUTOMATED LAB INSTRUMENTS;;Sep 27, 1994
3 ;CROSS LINK BY ID=ACCESSION OR IDE=MACHINE SEQUENCE OR TRAY/CUP
4LA1 S:$D(ZTQUEUED) ZTREQ="@" S LANM=$T(+0),TSK=$O(^LAB(62.4,"C",LANM,0)) Q:TSK<1
5 Q:'$D(^LA(TSK,"I",0))
6 K LATOP D ^LASET Q:'TSK S X="TRAP^"_LANM,@^%ZOSF("TRAP")
7LA2 K TV,Y S TOUT=0,A=2 D IN G QUIT:TOUT,LA2:CTRL'="C",LA2:"NEC"'[$E(IN,3) D QC
8 ;S TOUT=0,BAD=0 D IN,QC G QUIT:TOUT
9 F I=25:9:255 S V1=$E(Y(2),I,(I+8)),V=$E(V1,1,2) D NUM I V S TEST=+V,V=$E(V1,3,8) D NUM I V]"" S:$D(TC(TEST,1))#2 @TC(TEST,1)=V ;V COULD BE MODIFIED ON THIS LINE
10 I "NE"[$E(Y(2),3) S V=$E(Y(2),13,23) D NUM S ID=+V,V=$E(Y(2),4,7) D NUM S IDE=+V,TRAY=$E(Y(2),9),V=$E(Y(2),10,11) D NUM S CUP=+V
11 I $E(Y(2),3)="C" S V=$E(Y(2),6,7) D NUM S IDE=+V,(TRAY,CUP,ID)=""
12LA3 X LAGEN G LA2:'ISQN ;Can be changed by the cross-link code
13 F I=0:0 S I=$O(TV(I)) Q:I<1 S:TV(I,1)]"" ^LAH(LWL,1,ISQN,I)=TV(I,1)
14 G LA2
15QC ;QC TESTING HERE; S BAD=1 IF DONT STORE
16 S Y(A)=IN Q
17NUM S X="" F JJ=1:1:$L(V) S:$A(V,JJ)>32 X=X_$E(V,JJ)
18 S V=X Q
19IN S CNT=^LA(TSK,"I",0)+1 IF '$D(^(CNT)) S TOUT=TOUT+1 Q:TOUT>9 H 5 G IN
20 S ^LA(TSK,"I",0)=CNT,IN=^(CNT),TOUT=0
21 S:IN["~" CTRL=$P(IN,"~",2),IN=$P(IN,"~",1)
22 Q
23OUT S CNT=^LA(TSK,"O")+1,^("O")=CNT,^("O",CNT)=OUT
24 LOCK ^LA("Q") S Q=^LA("Q")+1,^("Q")=Q,^("Q",Q)=TSK LOCK
25 Q
26QUIT LOCK ^LA(TSK) H 1 K ^LA(TSK),^LA("LOCK",TSK),^TMP($J),^TMP("LA",$J)
27 Q
28TRAP D ^LABERR S T=TSK D SET^LAB G @("LA2^"_LANM) ;ERROR TRAP
Note: See TracBrowser for help on using the repository browser.