source: FOIAVistA/trunk/r/AUTOMATED_LAB_INSTRUMENTS-LA/LARA1K.m@ 1754

Last change on this file since 1754 was 628, checked in by George Lilly, 15 years ago

initial load of FOIAVistA 6/30/08 version

File size: 1.1 KB
Line 
1LARA1K ;SLC/RWF - TECHNICON RA-1000 ;7/20/90 10:02 ;
2 ;;5.2;AUTOMATED LAB INSTRUMENTS;;Sep 27, 1994
3 ;CROSS LINK BY ID OR IDE
4LA1 S:$D(ZTQUEUED) ZTREQ="@" S LANM=$T(+0),TSK=$O(^LAB(62.4,"C",LANM,0)) Q:TSK<1
5 Q:'$D(^LA(TSK,"I",0))
6 K LATOP D ^LASET Q:'TSK S X="TRAP^"_LANM,@^%ZOSF("TRAP")
7LA2 K TV,Y S TOUT=0,A=1 D IN G QUIT:TOUT,LA2:$E(IN,2)'="R"
8 S TOUT=0,BAD=0 D QC F A=2:1:3 D IN,QC G QUIT:TOUT
9 F I=1:13:239 S X1=$E(Y(2),I,I+13),II=+X1 I II,$D(TC(II)) S V=$E(X1,4,9) D NUM S:+V @TC(II,1)=+V
10 S V=$E(Y(1),4,10) D NUM S IDE=+V,ID=$E(Y(1),13,19),TRAY=+$P(ID,"-",1),CUP=+$P(ID,"-",2),ID=TRAY_CUP
11LA3 X LAGEN G LA2:'ISQN
12 F I=0:0 S I=$O(TV(I)) Q:I<1 S:TV(I,1)]"" ^LAH(LWL,1,ISQN,I)=TV(I,1)
13 G LA2
14QC ;QC TESTING HERE; S BAD=1 IF DONT STORE
15 S Y(A)=IN Q
16NUM S X="" F JJ=1:1:$L(V) S:$A(V,JJ)>32 X=X_$E(V,JJ)
17 S V=X Q
18IN S CNT=^LA(TSK,"I",0)+1 IF '$D(^(CNT)) S TOUT=TOUT+1 Q:TOUT>9 H 5 G IN
19 S ^LA(TSK,"I",0)=CNT,IN=^(CNT),TOUT=0
20 S:IN["~" CTRL=$P(IN,"~",2),IN=$P(IN,"~",1)
21 Q
22QUIT LOCK ^LA(TSK) H 1 K ^LA(TSK),^LA("LOCK",TSK),^TMP($J),^TMP("LA",$J)
23 Q
24TRAP D ^LABERR S T=TSK D SET^LAB G @("LA2^"_LANM) ;ERROR TRAP
Note: See TracBrowser for help on using the repository browser.